How can I customize my P300?
You can customize both the structure of the P300 and the built-in methods, but you will need to let us know your application or target compounds. We install particular columns and/or the detector at your request or base on the information you provided. We can also generate a built-in method with programming parameters of your target compounds all prese
How many compounds can be included in a built-in method?
For the default built-in method, the Air Quality method includes the most, with 21 target compounds. But with a customized built-in method, you can include as many compounds as you want for your application.
Is P300 able to detect both organic compounds and inorganic compounds?
Yes. Generally speaking, you can use it in any application as long as the organic compounds and inorganic gases can be separated and have ionization potential within 10.6 eV.
What's the detection limit of P300?
Sub-ppb (Benzene).
How do I prepare sample for P300?
If you want to analyze an air sample in a field, you don’t need any sample preparation at all. Simply point the sampling port into the area of your desired sample, and the P300 will pump in the air sample automatically when you click on “Start”.
If you want to analyze a gas sample in a container, attach the sampling needle onto the sampling port of the P300 and insert the needle into the container.
If you want to analyze a water sample in a field, you will need a purge and trap system prior to using the P300. Contact us for our PNT system.
How do I determine the sampling time?
For best results, the minimum sampling time is 6 seconds, or 0.1 minute. Here are some recommendations for sampling time based on estimated sample concentration: 5 minutes or more for 1 ppb or less, 2 ~ 5 minutes for 1 ppb ~ 500 ppb, 2 minutes or less for 1 ppm or above.
Do I need daily maintenance for P300?
No. The P300 will run auto cleaning after each test which saves you the maintenance work. But you can still run “Deep Clean” if you want to further clean the system.
Do I need a filter prior to using the P300?
Normally you won’t need a filter. But if there are large particles like dust or PM2.5 in the surrounding area, a filter is highly recommended.
When shall I charge my P300?
When only 2 out of 5 battery indication lights are on, you should charge the P300.
What carrier gas should I use? How pure should my carrier gas be?
Helium. 99.5% or purer.
Where can I attach the carrier gas?
There is a carrier gas cabin in the P300 for the installation of small disposable cylinders containing helium.
How long can the disposable carrier gas last?
Generally, in P300 system, a full cylinder of helium can support up to 80 hours of tests.
Can I travel to another city or another country with the P300?
Yes. The battery is permitted to be carried on board within the nation. However, a carrier gas cylinder is not permitted. You will need to vent and uninstall the gas cylinder before you check in at the airport.
What system should my computer have?
NovaSoft, the software to control the P300, will be installed on the built-in PC. To run the software on an additional computer, you will need a Win10 system.